Matta, Ragai-Edward, Bergauer, Bastian, Adler, Werner ORCID: 0000-0001-9791-5407, Wichmann, Manfred and Nickenig, Hans-Joachim (2017). The impact of the fabrication method on the three-dimensional accuracy of an implant surgery template. J. Cranio-MaxilloFac. Surg., 45 (6). S. 804 - 809. EDINBURGH: CHURCHILL LIVINGSTONE. ISSN 1878-4119
Full text not available from this repository.Abstract
Purpose: The use of a surgical template is a well-established method in advanced implantology. In addition to conventional fabrication, computer-aided design and computer-aided manufacturing (CAD/ CAM) work-flow provides an opportunity to engineer implant drilling templates via a three-dimensional printer. In order to transfer the virtual planning to the oral situation, a highly accurate surgical guide is needed. The aim of this study was to evaluate the impact of the fabrication method on the three-dimensional accuracy. Materials and methods: The same virtual planning based on a scanned plaster model was used to fabricate a conventional thermo4ormed and a three-dimensional printed surgical guide for each of 13 patients (single tooth implants). Both templates were acquired individually on the respective plaster model using an optical industrial white-light scanner (ATOS II, GOM mbh, Braunschweig, Germany), and the virtual datasets were superimposed. Using the three-dimensional geometry of the implant sleeve, the deviation between both surgical guides was evaluated. Results: The mean discrepancy of the angle was 3.479 (standard deviation, 1.904) based on data from 13 patients. Concerning the three-dimensional position of the implant sleeve, the highest deviation was in the Z-axis at 0.594 mm. The mean deviation of the Euclidian distance, dxyz, was 0.864 mm. Conclusion: Although the two different fabrication methods delivered statistically significantly different templates, the deviations ranged within a decimillimeter span. Both methods are appropriate for clinical use. (C) 2017 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.
Item Type: | Journal Article | ||||||||||||||||||||||||
Creators: |
|
||||||||||||||||||||||||
URN: | urn:nbn:de:hbz:38-230015 | ||||||||||||||||||||||||
DOI: | 10.1016/j.jcms.2017.02.015 | ||||||||||||||||||||||||
Journal or Publication Title: | J. Cranio-MaxilloFac. Surg. | ||||||||||||||||||||||||
Volume: | 45 | ||||||||||||||||||||||||
Number: | 6 | ||||||||||||||||||||||||
Page Range: | S. 804 - 809 | ||||||||||||||||||||||||
Date: | 2017 | ||||||||||||||||||||||||
Publisher: | CHURCHILL LIVINGSTONE | ||||||||||||||||||||||||
Place of Publication: | EDINBURGH | ||||||||||||||||||||||||
ISSN: | 1878-4119 | ||||||||||||||||||||||||
Language: | English | ||||||||||||||||||||||||
Faculty: | Unspecified | ||||||||||||||||||||||||
Divisions: | Unspecified | ||||||||||||||||||||||||
Subjects: | no entry | ||||||||||||||||||||||||
Uncontrolled Keywords: |
|
||||||||||||||||||||||||
Refereed: | Yes | ||||||||||||||||||||||||
URI: | http://kups.ub.uni-koeln.de/id/eprint/23001 |
Downloads
Downloads per month over past year
Altmetric
Export
Actions (login required)
View Item |