Goertz, Lukas, Liebig, Thomas, Siebert, Eberhard, Herzberg, Moriz, Neuschmelting, Hannah, Borggrefe, Jan ORCID: 0000-0003-2908-7560, Onur, Oezguer A., Dorn, Franziska and Kabbasch, Christoph (2020). Risk Factors of Procedural Complications Related to Woven EndoBridge (WEB) Embolization of Intracranial Aneurysms. Clin. Neuroradiol., 30 (2). S. 297 - 305. HEIDELBERG: SPRINGER HEIDELBERG. ISSN 1869-1447

Full text not available from this repository.

Abstract

Purpose The Woven EndoBridge (WEB) device has been proven to be a safe and efficient endovascular treatment option for wide-necked bifurcation aneurysms. The study aimed to evaluate the incidence and risk factors of procedural complications related to WEB embolization of ruptured and unruptured intracranial aneurysms. Methods This was a multicenter, observational study of consecutive patients with ruptured and unruptured aneurysms who were treated with the WEB at three German tertiary care centers between May 2011 and February 2018. Patient characteristics, anatomical details and procedural aspects were retrospectively collected and the impact on procedure-related complications was evaluated. Results Among 120 patients (mean age 58.5 +/- 11.9 years) with 120 aneurysms (mean size: 8.5 +/- 4.5 mm), WEB implantation was successful in 112 patients (93.3%). The rates for overall and symptomatic complications were 11.7% and 5.0%, respectively. At 6-month follow-up device-related morbidity was 1.2% among unruptured aneurysms and 2.6% among ruptured aneurysms. In the univariate analysis, a lower aspect ratio (p = 0.04) and an increased width-to-height ratio (p = 0.03) were significant risk factors for procedural complications. Conclusion The results of this study confirmed the WEB to be a safe treatment option, which is associated with low complication rates and minimal morbidity. Complications tended to occur more often in aneurysms with an unfavorable ratio between aneurysm height and aneurysm/neck width.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Goertz, LukasUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Liebig, ThomasUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Siebert, EberhardUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Herzberg, MorizUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Neuschmelting, HannahUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Borggrefe, JanUNSPECIFIEDorcid.org/0000-0003-2908-7560UNSPECIFIED
Onur, Oezguer A.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Dorn, FranziskaUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Kabbasch, ChristophUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-331395
DOI: 10.1007/s00062-019-00762-8
Journal or Publication Title: Clin. Neuroradiol.
Volume: 30
Number: 2
Page Range: S. 297 - 305
Date: 2020
Publisher: SPRINGER HEIDELBERG
Place of Publication: HEIDELBERG
ISSN: 1869-1447
Language: English
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
SAFETY; COILINGMultiple languages
Clinical Neurology; Radiology, Nuclear Medicine & Medical ImagingMultiple languages
URI: http://kups.ub.uni-koeln.de/id/eprint/33139

Downloads