Oikonomidis, Stavros, Beyer, Frank, Meyer, Carolin, Baltin, Christoph Tobias, Eysel, Peer and Bredow, Jan (2020). Insertion Angle of Pedicle Screws in the Subaxial Cervical Spine: The Analysis of Computed Tomography-Navigated Insertion of Pedicle Screws. Asian Spine J., 14 (1). S. 66 - 72. SEOUL: KOREAN SOC SPINE SURGERY. ISSN 1976-7846

Full text not available from this repository.

Abstract

Study Design: Four orthopedic spine surgeons measured the radiological parameters of pedicle screws in the cervical spine using a postoperative computed tomography (CT) scan. Purpose: This study analyzed the insertion angle of CT-navigated insertion of pedicle screws in the subaxial cervical spine and classified them according to their position. Overview of Literature: Overall, a pedicle transverse angle of 33.6 degrees-50.2 degrees with a mean angle of 45 degrees relative to the midline has been reported in the literature. Methods: The insertion angles of 87 pedicle screws inserted using CT-based navigation in the subaxial cervical spine were measured in the postoperative CT. The screw positioning was determined according to the modified Gertzbein and Robbins classification. Results: Total 89.3% (n=78) of the pedicle screws inserted using CT-based navigation showed good placement. The mean insertion angle of the pedicle screws that showed good positioning was 29.9 degrees +/- 9.9 degrees. The pedicle screws showing bad positioning had a mean insertion angle of 26.8 degrees +/- 10.5 degrees (p=0.157). The interobserver reliability showed a reliable measurement intraclass correlation coefficient: 0.994 (95% confidence interval, 0.992-0.996). Conclusions: The present results show that the insertion angle of the pedicle screws in the subaxial cervical spine was smaller than the actual pedicle transverse angle, as per the literature. One reason for this discrepancy could be that the navigation systems allow the insertion of cervical pedicle screws with a lower convergence.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Oikonomidis, StavrosUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Beyer, FrankUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Meyer, CarolinUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Baltin, Christoph TobiasUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Eysel, PeerUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Bredow, JanUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-347252
DOI: 10.31616/asj.2019.0053
Journal or Publication Title: Asian Spine J.
Volume: 14
Number: 1
Page Range: S. 66 - 72
Date: 2020
Publisher: KOREAN SOC SPINE SURGERY
Place of Publication: SEOUL
ISSN: 1976-7846
Language: English
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
ANATOMICAL CONSIDERATIONS; FIXATION; ACCURACY; FEASIBILITY; PLACEMENT; SURGERY; LESIONSMultiple languages
OrthopedicsMultiple languages
URI: http://kups.ub.uni-koeln.de/id/eprint/34725

Downloads

Downloads per month over past year

Altmetric

Export

Actions (login required)

View Item View Item