Rosentreter, A., Hoerster, R., Schick, T., Eter, N., Dietlein, T. S. and Fauser, S. (2015). Rebound tonometry after vitreoretinal surgery. Ophthalmologe, 112 (11). S. 917 - 923. NEW YORK: SPRINGER. ISSN 1433-0423

Full text not available from this repository.

Abstract

The aim of this study was to analyze the practicability and comparability of the Icare rebound tonometer (RT) versus the Schiotz indentation tonometer (SIT) and the Goldmann applanation tonometer (GAT) for measuring intraocular pressure (IOP) in patients after pars plana vitrectomy (PPV). A total of 100 eyes from 100 patients who underwent vitreoretinal surgery in the Department of Ophthalmology, University of Cologne were included in this prospective analysis. The IOP was measured using RT preoperatively, on the day of surgery and 2 days after surgery, using SIT on the day of surgery and GAT preoperatively and 2 days after surgery. For the evaluation eyes were divided into subgroups with respect to the endotamponade selected and the IOP level. The mean preoperative IOP for all enrolled eyes was 15.4 +/- 8.0 mmHg for RT and 16.1 +/- 7.9 mmHg for GAT. Bland-Altman analysis revealed a bias between RT and GAT of -aEuro parts per thousand 0.6 mmHg. Bland-Altman analysis for the postoperative course of all eyes revealed a bias of 3.0 mmHg between RT and SIT on the day of surgery and no bias between RT and GAT in the further postoperative follow-up. Rebound tonometry seems to provide precise IOP values after vitreoretinal surgery. Divergence from SIT values on the day of surgery is presumably due to a general tendency of SIT to underestimate IOP values. Therefore, RT can be used in the clinical routine after vitreoretinal surgery as an alternative to GAT.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Rosentreter, A.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Hoerster, R.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Schick, T.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Eter, N.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Dietlein, T. S.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Fauser, S.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-387960
DOI: 10.1007/s00347-015-0076-9
Journal or Publication Title: Ophthalmologe
Volume: 112
Number: 11
Page Range: S. 917 - 923
Date: 2015
Publisher: SPRINGER
Place of Publication: NEW YORK
ISSN: 1433-0423
Language: German
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
GOLDMANN APPLANATION TONOMETER; CENTRAL CORNEAL THICKNESS; INTRAOCULAR-PRESSURE; ICARE; VITRECTOMYMultiple languages
OphthalmologyMultiple languages