Wichtmann, Hildegard M., Laukamp, Kai R., Manneck, Sebastian, Appelt, Konrad, Stieltjes, Bram, Boll, Daniel T., Benz, Matthias R. and Obmann, Markus M. (2023). Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction? Abdom. Radiol., 48 (1). S. 424 - 436. NEW YORK: SPRINGER. ISSN 2366-0058

Full text not available from this repository.

Abstract

Purpose To assess image quality and metal artifact reduction in split-filter dual-energy CT (sfDECT) of the abdomen with hip or spinal implants using virtual monoenergetic images (VMI) and iterative metal artifact reduction algorithm (iMAR). Methods 102 portal-venous abdominal sfDECTs of patients with hip (n = 71) or spinal implants (n = 31) were included in this study. Images were reconstructed as 120kVp-equivalent images (Mixed) and VMI (40-190 keV), with and without iMAR. Quantitative artifact and image noise was measured using 12 different ROIs. Subjective image quality was rated by two readers using a five- point Likert-scale in six categories, including overall image quality and vascular contrast. Results Lowest quantitative artifact in both hip and spinal implants was measured in VMI190keV-iMAR. However, it was not significantly lower than in Mixed(iMAR) (for all ROIs, p = 1.00), which were rated best for overall image quality (hip: 1.00 [IQR: 1.00-2.00], spine: 3.00 [IQR:2.00-3.00]). VMI50keV- iMAR was rated best for vascular contrast (hip: 1.00 [IQR: 1.00-2.00], spine: 2.00 [IQR: 1.00-2.00]), which was significantly better than Mixed (both, p < 0.001). VMI50keV-iMAR provided superior overall image quality compared to Mixed for hip (1.00 vs 2.00, p < 0.001) and similar diagnostic image quality for spinal implants (2.00 vs 2.00, p = 0.51). Conclusion For abdominal sfDECT with hip or spinal implants Mixed(iMAR) images should be used. High keV VMI do not further improve image quality. IMAR allows the use of low keV images (VMI50keV) to improve vascular contrast, compared to Mixed images.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Wichtmann, Hildegard M.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Laukamp, Kai R.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Manneck, SebastianUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Appelt, KonradUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Stieltjes, BramUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Boll, Daniel T.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Benz, Matthias R.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Obmann, Markus M.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-675938
DOI: 10.1007/s00261-022-03682-3
Journal or Publication Title: Abdom. Radiol.
Volume: 48
Number: 1
Page Range: S. 424 - 436
Date: 2023
Publisher: SPRINGER
Place of Publication: NEW YORK
ISSN: 2366-0058
Language: English
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
COMPUTED BODY TOMOGRAPHY; MAGNETIC-RESONANCE; REDUCING ARTIFACTS; IMAGE-QUALITY; WHITE PAPER; HARDWARE; RECONSTRUCTION; SOFTWARE; SOCIETY; SPINEMultiple languages
Radiology, Nuclear Medicine & Medical ImagingMultiple languages
URI: http://kups.ub.uni-koeln.de/id/eprint/67593

Downloads

Downloads per month over past year

Altmetric

Export

Actions (login required)

View Item View Item