Kuffer, Liz, Steven, Daniel, Halbach, Marcel, Lueker, Jakob, van den Bruck, Jan-Hendrik and Sultan, Arian (2017). Combination of a subcutaneous ICD in a patient with a baroreceptor activation device: Feasibility, safety, and precautions: A Case Report. PACE-Pacing Clin. Electrophysiol., 40 (12). S. 1486 - 1489. HOBOKEN: WILEY. ISSN 1540-8159

Full text not available from this repository.

Abstract

We present a case of a patient with a baroreflex activation therapy (BAT) receiving a subcutaneous implantable cardioverter defibrillator (S-ICD). We anticipated two possible hazardous interactions between the two devices. Stimulation by the BAT could be adjudicated as noise and result in underdetection of ventricular arrhythmias or it might be misinterpreted as ventricular arrhythmias and lead to inappropriate shocks. Postop ensing occurred, the upper limit of pulse width of the BAT was limited because of noise detection by the S-ICD, but the upper limit of amplitude was limited by patient's discomfort. In this patient, the combination of a BAT and an S-ICD was safe.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Kuffer, LizUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Steven, DanielUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Halbach, MarcelUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Lueker, JakobUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
van den Bruck, Jan-HendrikUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Sultan, ArianUNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-209076
DOI: 10.1111/pace.13109
Journal or Publication Title: PACE-Pacing Clin. Electrophysiol.
Volume: 40
Number: 12
Page Range: S. 1486 - 1489
Date: 2017
Publisher: WILEY
Place of Publication: HOBOKEN
ISSN: 1540-8159
Language: English
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
IMPLANTABLE-CARDIOVERTER-DEFIBRILLATOR; RESISTANT HYPERTENSION; HEART-FAILURE; THERAPYMultiple languages
Cardiac & Cardiovascular Systems; Engineering, BiomedicalMultiple languages
Refereed: Yes
URI: http://kups.ub.uni-koeln.de/id/eprint/20907

Downloads

Downloads per month over past year

Altmetric

Export

Actions (login required)

View Item View Item