Hose, J. ORCID: 0000-0003-4085-2219, Juebner, M. ORCID: 0000-0002-5145-4884, Thevis, M. and Andresen-Streichert, H. ORCID: 0000-0002-8779-8902 (2025). Assessment and cross-validation of calibration transferability between dried blood spot sampling devices for accurate quantification of phosphatidylethanol. Analytical and Bioanalytical Chemistry, 417 (17). pp. 3803-3811. Springer Nature. ISSN 1618-2642

[thumbnail of s00216-025-05897-x.pdf] PDF
s00216-025-05897-x.pdf
Bereitstellung unter der CC-Lizenz: Creative Commons Attribution.

Download (2MB)
Identification Number:10.1007/s00216-025-05897-x

Abstract

Phosphatidylethanol (PEth) is a direct alcohol biomarker which is used in a variety of clinical and forensic contexts. Due to its limited stability in whole blood, PEth is routinely analysed using dried blood spots (DBS). There are many different commercially available sampling systems for DBS, but data on their comparability is scarce. The aim of this study was to develop and validate a LC-MS/MS method for the quantification of PEth 16:0/18:1 and PEth 16:0/18:2 for three different DBS sampling systems: Whatman™ filter paper (903 Protein Saver Cards), Mitra ® and Capitainer ® B Vanadate. The transferability of calibration curves and their applicability across these devices should be assessed. The results showed that DBS of all devices could be prepared and analysed using the same procedure. Validation was successful for the devices. The accuracy of QCs was best if quantified using the matching DBS sampling device calibration curve (bias < 15 %). Higher, but still acceptable, bias values were observed when using a Whatman™ calibration curve to quantify PEth in DBS of the Capitainer ® B Vanadate system and vice versa. Mitra ® QCs should only be quantified using device-specific calibration curves, as in all other cases, accuracy exceeded the limit of 15 %. We recommend analysing samples using device-specific calibrations at all times, as determined PEth concentrations might be crucial for patients and should not be subjected to any avoidable uncertainty.

Item Type: Article
Creators:
Creators
Email
ORCID
ORCID Put Code
Hose, J.
UNSPECIFIED
UNSPECIFIED
Juebner, M.
UNSPECIFIED
UNSPECIFIED
Thevis, M.
UNSPECIFIED
UNSPECIFIED
UNSPECIFIED
Andresen-Streichert, H.
UNSPECIFIED
UNSPECIFIED
URN: urn:nbn:de:hbz:38-801026
Identification Number: 10.1007/s00216-025-05897-x
Journal or Publication Title: Analytical and Bioanalytical Chemistry
Volume: 417
Number: 17
Page Range: pp. 3803-3811
Number of Pages: 9
Date: 15 July 2025
Publisher: Springer Nature
ISSN: 1618-2642
Language: English
Faculty: Faculty of Medicine
Divisions: Faculty of Medicine > Rechtsmedizin > Institut für Rechtsmedizin
Subjects: Medical sciences Medicine
Uncontrolled Keywords:
Keywords
Language
Alcohol biomarker ; Alternative matrices ; Sampling techniques ; LC-MS/MS
English