Goertz, Lukas ORCID: 0000-0002-2620-7611, Al-Sewaidi, Yosef, Habib, Mahmoud, Zopfs, David ORCID: 0000-0001-9978-7453, Reichardt, Benjamin, Ranft, Alexander and Kabbasch, Christoph ORCID: 0000-0003-3712-2258 (2024). State-of-the-art mobile head CT scanner delivers nearly the same image quality as a conventional stationary CT scanner. Scientific Reports, 14 (1). pp. 1-9. Springer Nature. ISSN 2045-2322

[thumbnail of s41598-024-56089-z.pdf] PDF
s41598-024-56089-z.pdf
Bereitstellung unter der CC-Lizenz: Creative Commons Attribution.

Download (1MB)
Identification Number:10.1038/s41598-024-56089-z

Abstract

[Artikel-Nr.: 6393] The use of mobile head CT scanners in the neurointensive care unit (NICU) saves time for patients and NICU staff and can reduce transport‑related mishaps, but the reduced image quality of previous mobile scanners has prevented their widespread clinical use. This study compares the image quality of SOMATOM On.Site (Siemens Healthineers, Erlangen, Germany), a state‑of‑the‑art mobile head CT scanner, and a conventional 64‑slice stationary CT scanner. The study included 40 patients who underwent head scans with both mobile and stationary scanners. Gray and white matter signal and noise were measured at predefined locations on axial slices, and signal‑to‑noise ratios (SNRs) and contrast‑to‑noise ratios (CNRs) were calculated. Artifacts below the cranial calvaria and in the posterior fossa were also measured. In addition, image quality was subjectively assessed by two radiologists in terms of corticomedullary differentiation, subcalvarial space, skull artifacts, and image noise. Quantitative measurements showed significantly higher image quality of the stationary CT scanner in terms of noise, SNR and CNR of gray and white matter. Artifacts measured in the posterior fossa were higher with the mobile CT scanner, but subcalvarial artifacts were comparable. Subjective image quality was rated similarly by two radiologists for both scanners in all domains except image noise, which was better for stationary CT scans. The image quality of the SOMATOM On.Site for brain scans is inferior to that of the conventional stationary scanner, but appears to be adequate for daily use in a clinical setting based on subjective ratings.

Item Type: Article
Creators:
Creators
Email
ORCID
ORCID Put Code
Goertz, Lukas
UNSPECIFIED
UNSPECIFIED
Al-Sewaidi, Yosef
UNSPECIFIED
UNSPECIFIED
UNSPECIFIED
Habib, Mahmoud
UNSPECIFIED
UNSPECIFIED
UNSPECIFIED
Zopfs, David
UNSPECIFIED
UNSPECIFIED
Reichardt, Benjamin
UNSPECIFIED
UNSPECIFIED
UNSPECIFIED
Ranft, Alexander
UNSPECIFIED
UNSPECIFIED
UNSPECIFIED
Kabbasch, Christoph
UNSPECIFIED
UNSPECIFIED
URN: urn:nbn:de:hbz:38-815270
Identification Number: 10.1038/s41598-024-56089-z
Journal or Publication Title: Scientific Reports
Volume: 14
Number: 1
Page Range: pp. 1-9
Number of Pages: 9
Date: 16 March 2024
Publisher: Springer Nature
ISSN: 2045-2322
Language: English
Faculty: Faculty of Medicine
Divisions: Faculty of Medicine > Radiologische Diagnostik > Institut und Poliklinik für Radiologische Diagnostik
Subjects: Medical sciences Medicine
Uncontrolled Keywords:
Keywords
Language
Computed tomography ; ICU ; Intracranial haemorrhage ; Mobile CT
English
['eprint_fieldname_oa_funders' not defined]: Publikationsfonds UzK
Refereed: Yes
URI: http://kups.ub.uni-koeln.de/id/eprint/81527

Downloads

Downloads per month over past year

Altmetric

Export

Actions (login required)

View Item View Item